Seed BOD calculation

Mastering BOD Analysis: The Essentials of Seeding and Estimation

Biochemical Oxygen Demand (BOD) is a cornerstone parameter in water and wastewater analysis. It indicates how much dissolved oxygen microorganisms require to biologically degrade organic matter present in a sample. Accurate BOD results are essential for assessing pollution levels, evaluating treatment efficiency, and protecting receiving waters.

However, reliable BOD testing depends on two often‑misunderstood steps: proper seeding and realistic BOD estimation before dilution. This blog walks through the why, when, and how of seeding, explains practical BOD estimation techniques, and shows you how to select the right sample volume to avoid failed tests.




Why Seeding Matters in BOD Analysis

BOD testing is fundamentally a biological process. Microorganisms consume biodegradable organic matter and, in doing so, deplete dissolved oxygen. If sufficient and active microbes are not present, oxygen consumption will be low—even when organic pollution is high—leading to falsely low BOD values.

When Seeding Is Required

Seeding is necessary when the sample does not contain enough viable microorganisms, such as:

  • Industrial wastewater (especially toxic or disinfected streams)
  • Treated effluents after chlorination or advanced treatment
  • Samples stored for extended periods

In these cases, a microbial seed (commonly from domestic wastewater, activated sludge, or commercial seed preparations) is added to ensure biological oxidation can proceed.

When Seeding Is Not Required

Seeding is usually unnecessary for samples already rich in microorganisms, including:

  • Raw sewage
  • River or stream water receiving wastewater discharges

These samples naturally contain adequate bacterial populations to carry out biodegradation.

Proper seeding ensures the BOD test simulates natural conditions and produces representative, defensible results.


Estimating BOD Before Testing

One of the most common causes of BOD test failure is incorrect sample dilution. If too much sample is used, dissolved oxygen may drop to zero before the test ends. If too little is used, the oxygen depletion may be too small to measure accurately.

To avoid this, always estimate BOD before setting up dilutions.

Using COD as a BOD Estimation Tool

Chemical Oxygen Demand (COD) is a fast, chemical measurement of oxidizable material and is often used as a guide for BOD estimation.

A widely used rule of thumb:

Estimated BOD ≈ 70% of COD

This accounts for the fact that not all oxidizable material measured by COD is biologically degradable.

This estimate allows you to select dilutions that will result in an oxygen depletion of 2–7 mg/L, the optimal range for valid BOD results.


Calculating the Correct Sample Volume

Standard BOD bottles hold 300 mL of diluted sample. The following formula is used to calculate the sample volume required per liter of dilution water:

Sample volume (mL/L) = (X ÷ Estimated BOD) × 1000

Where X is the target oxygen depletion (mg/L).

Choosing X Values

  • Two dilutions (recommended): X = 2.5 and 4.0 mg/L
  • Single dilution: X = 3.0 or 3.5 mg/L

Always round to a practical volume for pipetting. For samples with very high BOD, perform a pre‑dilution with distilled water before preparing final BOD bottles.


Worked Example

Given: COD = 400 mg/L

Step 1: Estimate BOD

Estimated BOD = 400 × 0.7 = 280 mg/L

Step 2: Calculate sample volumes

  • X = 2.5 mg/L:

    • (2.5 ÷ 280) × 1000 = 8.9 mL/L (≈ 9 mL/L)
    • For a 300 mL bottle: 9 × (300 ÷ 1000) = 2.7 mL
  • X = 4.0 mg/L:

    • (4.0 ÷ 280) × 1000 = 14.3 mL/L (≈ 14 mL/L)
    • For a 300 mL bottle: 14 × (300 ÷ 1000) = 4.2 mL

Add the calculated sample volume to each BOD bottle, fill with aerated dilution water containing nutrients and seed (if required), and incubate at 20°C for 5 days. Measure dissolved oxygen before and after incubation to determine BOD.


Practical Tips and Common Pitfalls

  • Use high‑quality dilution water: Free from chlorine, toxic metals, and inhibitors
  • Seed controls are essential: Always run seed blanks to correct for seed oxygen uptake
  • Watch for toxicity: No oxygen depletion may indicate inhibitory substances
  • Avoid overloading: Zero DO invalidates the test
  • Always estimate first: Skipping estimation is the fastest way to fail a BOD test

Final Thoughts

Seeding and BOD estimation are not optional extras—they are fundamental to producing valid and meaningful BOD data. By estimating BOD from COD, selecting proper dilutions, and applying seeding only when necessary, you improve accuracy, save time, and reduce repeat testing.

Accurate BOD measurements support better environmental decisions, effective wastewater treatment, and healthier water bodies.

If you’ve applied these methods in your lab, share your experience or questions. For deeper dives, explore related topics such as COD testing methods, seed correction calculations, or microbial inhibition screening.

Happy testing—and keep your oxygen in the sweet spot!