Skip to main content

Biological Oxygen Demand BOD Teating and Measurement in water and waste water

 PURPOSE:

To describe the Laboratory standard operating procedure for measurement of Biological Oxygen Demand of Water and Waste Water samples.


SCOPE:

Applicable to the Laboratory for analysis of Water and Waste Water samples, where measurement of Biological oxygen demand is required.

 

RESPONSIBILITY: 

Lab chemist: Analysis of sample for measurement of Biological Oxygen demand.

Technical Manager:   Review of activity 

Quality Manager: Implementation and compliance of SOP.

PROCEDURE: 

Principle:

Biochemical Oxygen Demand (BOD) is defined as the amount of O2 required by microorganisms while stabilizing biologically decomposable organic matter in a waste under aerobic conditions.



Instrument and Equipment:

BOD bottles 300ml capacity, BOD Incubator to be controlled at 27oC ± 1oC.

Reagents: 

Phosphate buffer: Dissolve 8.5 g KH2PO4, 21.75 g K2HPO4, 33.4 g Na2HPO4.7H2O and 1.7 g NH4Cl in distilled water and dilute to 1000ml. Adjust its pH to 7.2.

Magnesium sulfate: Dissolve 82.5 g MgSO4.7H2O and dilute it to 1000ml.

Calcium chloride: Dissolve 27.5 g CaCl2 and dilute it to 1000ml.

Ferric chloride solution: Dissolve 0.25 g FeCl3.6H2O and dilute it to 1000ml.

Sodium thiosulfate solution 0.025 N: Dissolve 6.205 g Na2S2O3 and dilute to 1000ml.

Test Method:

Preparation of dilution water:

Aerate the required volume of distilled water in a container by bubbling compressed air for 1-2 days to attain DO saturation. 

Add 1ml each of phosphate buffer, magnesium sulfate, calcium chloride and ferric chloride solution for each liter of dilution water. Mix well.

In the case of the wastes which are not expected to have sufficient bacterial population add seed to the dilution water. Generally, 2ml settled sewage is considered sufficient for 1000ml of dilution water

Dilution of sample:

Neutralize the sample to pH amount 7.0 if it is highly alkaline or acidic.

The sample should free from residual chlorine. If it contains residual chlorine   remove it by using Na2S2O3 solution as follows.

Removal of residual chlorine:

Take 50ml sample and acidify with 10ml of 1 + 1 acetic acid.

Add approximately 1g of KI

Titrate with Na2S2O3 using starch as indicator.

Calculate the amount of Na2S2O3 required per ml of the sample and add accordingly to the sample tested for BOD.

Samples having high DO content, i.e., DO 9 mg/L due to either algal growth or some other reason, reduce the DO content by aerating the agitating the samples.

Make several dilutions of the prepared samples so as to obtain about 50% depletion of DO in the dilution water but no less than 2 mg and the residual oxygen after 3 days of incubation should not be less than 1mg/L.Prepare dilution as follows:

Siphon out seeded dilution water in a measuring cylinder or volumetric flak half the required volume. Add the required quantity of carefully mixed sample. Dilute to the desired volume by siphoning dilution water and mix well.


The following dilutions are suggested:

  %   dilution                 BOD Range

      0.01                         20,000 -70,000

      0.02                          10,000 - 35,000

      0.05                          4000 -  14,000

       0.1                           2000 - 7000

       0.2                           1000 - 3500

       0.5                            400 - 1900

         1                              200 - 700

         2                              100 - 350

        5                               40 - 140

       10                               20 - 70

        20                              10 - 35

        50                               Up to 14

        100                             1 - 7


Siphon the dilution prepared as above labeled BOD bottles as demonstrated and stopper immediately.

Keep 1 bottle for determination of the initial DO and incubate 3 bottles at 27oC for 3 days. See that the bottle has a water seal.

Prepare a blank in duplicate by siphoning plain dilution water (without seed) to measure the O2 consumption in dilution water.

Fix the bottles kept for immediate DO determination and blank by adding 2ml MnSO4 followed by 2ml NaOH + KI + NaN3 as described in the measurement of DO

Determine DO in the sample and in the blank on initial day and after 3 days


Calculation:

Calculate BOD of the ample as follows

Let D0    =    DO in the sample bottle on 0th day.

               D1    =    DO in the sample bottle on 3rd day.

                   C0    =    DO in the blank bottle on 0th day.

                   C1    =    DO in the blank bottle on 3rd day   

 C0 - C1 =     DO depletion in the dilution water alone

 D0 - D1   =    DO depletion in the sample + dilution water

 (D0 -   D1) -  (C0 - C1) = DO depletion due to microbes

 BOD mg/L = (D0 - D1) – (C0 – C1) mg X decimal fraction of sample used.

If the sample is seeded find out BOD of seed in the above manner and apply corrections, as per demonstrated.


Comments

Popular posts from this blog

TOC analyser principle

1. A Total Organic Carbon (TOC) analyzer measures the amount of carbon in organic compounds present in a sample. The principle behind a TOC analyzer is based on the combustion or oxidation of organic materials to convert carbon into carbon dioxide (CO2), which is then measured to determine the total amount of organic carbon in the sample. There are two main methods for measuring TOC: Non-Combustion Method (Oxidation Method) : The sample is oxidized, usually using a strong oxidizing agent like persulfate or UV light, to convert organic carbon into CO2. The resulting CO2 is then measured using a non-dispersive infrared (NDIR) detector or another detection method to quantify the total organic carbon. Combustion Method : The sample is heated to high temperatures (around 680–950°C), where organic carbon is combusted in the presence of oxygen to form CO2. The CO2 produced is then measured using infrared spectroscopy or other appropriate detection techniques. Both methods al...

TDS- Total Dissolve Solid

Total Dissolved Solids (TDS) testing for wastewater is a critical procedure outlined in the Standard Methods for the Examination of Water and Wastewater by the American Public Health Association (APHA). Specifically, the method referenced is APHA Standard Method 2540C: Total Dissolved Solids Dried at 180°C. Below is a detailed Standard Operating Procedure (SOP) for preparing and conducting TDS testing in wastewater, including a description of the method and a flow chart to illustrate the process. Standard Operating Procedure (SOP) for TDS Testing in Wastewater (APHA Method 2540C) 1. Purpose To determine the concentration of total dissolved solids (TDS) in wastewater samples using the gravimetric method as specified in APHA Standard Method 2540C. TDS measures the mass of organic and inorganic substances dissolved in water that pass through a filter, typically after evaporation at 180°C. 2. Scope This SOP applies to the analysis of wastewater samples (industrial, municipal, or environmen...

Standard Operating Procedure for the Measurement of Total Hardness

  SOP For Total Hardness in Water and waste water EDTA Titrimetric method Purpose: The purpose of this protocol is to provide guidelines for analysis of Total Hardness in Water. Principle of method: EDTA and it's sodium salts from a chelated solution complex when added to a solution of certain metal actions, if a small amount of dye such as Eriochrome Black T is added to an aqoeous solution containing calcium and magnesium ions at a pH of 10+0.1 to 10-0.1, the solution becomes winr red. When titrated with EDTA , the magnesium and calcium will be complexed, and when all of the  and  has been complexed the solution turns from wine red to Blue.      To   Instrument/Equipment: The following items are necessary to perform the analysis of Total Hardness in Water; pH Meter Burette – 50 ml capacity Glass Wares Reagents/Chemicals: All the chemicals should be of Analytical Reagent Grade. Distilled Water Buffer Solution – Dissolve 16.9 g of Ammonium Chloride...