Standard Operating Prucedure for the Measurement of Chemicsl Oxygen Demand

 PURPOSE:

To describe the Laboratory standard operating procedure for measurement of Chemical Oxygen Demand of Water and Waste Water samples.



SCOPE:

Applicable to the Laboratory for analysis of Water and Waste Water samples, where measurement of Chemical Oxygen Demand is required.

 

RESPONSIBILITY: 

Lab chemist:

Analysis of sample for measurement of Chemical Oxygen Demand.

Technical Manager:

Review of activity 

Quality Manager:

Implementation and compliance of SOP.


PROCEDURE: 

Principle:

A sample is refluxed in strong acidic solution with a known excess of potassium dichromate (K2Cr2O7).After digestion the remaining unreduced K2Cr2O7 is titrated with ferrous ammonium sulphate to determine the amount of potassium dichromate (K2Cr2O7) consumed and the oxidizable matter is calculated in terms of oxygen equivalent.

 Instrument and Equipment:

Heating Mantle, Burette 50 ml.

Round Bottom Flask 500 ml

Pipettes 50ml.

Measuring Cylinders

Reagents: 

Preparation of Potassium Dichromate Solution, 0.25N: Dissolve 12.259g K2Cr2O7 in distilled water and dilute it to 1000ml.

Ferrous Ammonium Sulphate, 0.25 N: Dissolve 98 g Fe(NH4)2(SO4)2.6H2O in distilled water. Add 20ml conc. H2SO4, cool, and dilute to 1000ml.Standardize this solution against K2Cr2O7 daily.

Ferroin indicator : you can purchase of Ferroin indicator from the market or prepare in your laboratory like that Dissolve 1.485g 1, 10-phenanthroline monohydrate and 695mg FeSO4.7H2O in distilled water and dilute to 100ml. 

Sulphuric Acid Conc.

Silver Sulphate

Mercuric Sulphate

Test Method:

Pipette out 20ml. Sample into a 500 ml. round bottom refluxing flask 

Note: For higher COD value, use smaller quantity of sample diluted to 20ml with distilled water, so that the color of the reaction mixture will remain yellow to greenish yellow after adding reagents and chemicals mentioned.

Add app 0.5 gm of HgSO4 and a pinch of Ag2SO4 to it.

Add 10ml of 0.25 N K2Cr2O7 solutions to it.

Add 30 ml of conc. H2SO4 with gentle swirling of reaction mixture.

Reflux the reaction mixture on a heating mantle for minimum two hours with a             water condenser attached to it.

Cool and wash down condenser with about 80ml of distilled water.

 Cool to room temperature and titrate the excess K2Cr2O7 with 0.25 N Ferrous Ammonium Sulphate solution, using 4 to 5 drops of Ferroin indicator.

End Point of titration is judged by color change from Bluish green to Reddish Brown.

Note down the Burette Reading (Sample Reading).

Carry out Blank determination under similar conditions with 20ml-distilled water, using same quantity of reagents and chemicals. (Blank Reading)

5. Calculation:

COD (mg/L)     =    (B-A) X Nor of FAS X 8000   

                                                   Sample taken in ml



Where,   B = Volume of FAS required for blank

                A = Volume of FAS required for sample.



Comments

Post a Comment

Popular posts from this blog

TOC analyser principle

1. A Total Organic Carbon (TOC) analyzer measures the amount of carbon in organic compounds present in a sample. The principle behind a TOC analyzer is based on the combustion or oxidation of organic materials to convert carbon into carbon dioxide (CO2), which is then measured to determine the total amount of organic carbon in the sample. There are two main methods for measuring TOC: Non-Combustion Method (Oxidation Method) : The sample is oxidized, usually using a strong oxidizing agent like persulfate or UV light, to convert organic carbon into CO2. The resulting CO2 is then measured using a non-dispersive infrared (NDIR) detector or another detection method to quantify the total organic carbon. Combustion Method : The sample is heated to high temperatures (around 680–950°C), where organic carbon is combusted in the presence of oxygen to form CO2. The CO2 produced is then measured using infrared spectroscopy or other appropriate detection techniques. Both methods al...

TDS- Total Dissolve Solid

Total Dissolved Solids (TDS) testing for wastewater is a critical procedure outlined in the Standard Methods for the Examination of Water and Wastewater by the American Public Health Association (APHA). Specifically, the method referenced is APHA Standard Method 2540C: Total Dissolved Solids Dried at 180°C. Below is a detailed Standard Operating Procedure (SOP) for preparing and conducting TDS testing in wastewater, including a description of the method and a flow chart to illustrate the process. Standard Operating Procedure (SOP) for TDS Testing in Wastewater (APHA Method 2540C) 1. Purpose To determine the concentration of total dissolved solids (TDS) in wastewater samples using the gravimetric method as specified in APHA Standard Method 2540C. TDS measures the mass of organic and inorganic substances dissolved in water that pass through a filter, typically after evaporation at 180°C. 2. Scope This SOP applies to the analysis of wastewater samples (industrial, municipal, or environmen...

Standard Operating Procedure for the Measurement of Total Hardness

  SOP For Total Hardness in Water and waste water EDTA Titrimetric method Purpose: The purpose of this protocol is to provide guidelines for analysis of Total Hardness in Water. Principle of method: EDTA and it's sodium salts from a chelated solution complex when added to a solution of certain metal actions, if a small amount of dye such as Eriochrome Black T is added to an aqoeous solution containing calcium and magnesium ions at a pH of 10+0.1 to 10-0.1, the solution becomes winr red. When titrated with EDTA , the magnesium and calcium will be complexed, and when all of the  and  has been complexed the solution turns from wine red to Blue.      To   Instrument/Equipment: The following items are necessary to perform the analysis of Total Hardness in Water; pH Meter Burette – 50 ml capacity Glass Wares Reagents/Chemicals: All the chemicals should be of Analytical Reagent Grade. Distilled Water Buffer Solution – Dissolve 16.9 g of Ammonium Chloride...