Skip to main content

Toc

 Below is an updated **Standard Operating Procedure (SOP)** for **Total Organic Carbon (TOC)** analysis in **solid samples** using a TOC analyzer with a solid sample module (e.g., Shimadzu TOC-L with SSM-5000A). This SOP incorporates a detailed **calibration procedure** with specific criteria, establishes **criteria for quality control (QC) samples**, and includes the **computation for obtaining TOC** (TOC = TC - IC). The procedure is tailored for solid samples (e.g., soil, sediment, sludge) and includes an example for clarity.


---


### Standard Operating Procedure (SOP) for TOC Analysis of Solid Samples


#### 1. Purpose

To provide a standardized procedure for determining Total Organic Carbon (TOC) in solid samples using a TOC analyzer with a solid sample module, ensuring accurate, reproducible, and compliant results.


#### 2. Scope

This SOP applies to the analysis of TOC in solid samples (e.g., soil, sediment, sludge, or biomass). It covers sample preparation, instrument calibration, quality control, TOC computation, and data reporting.


#### 3. References

- ASTM D7573: Standard Test Method for TOC in Solid Samples

- EPA Method 9060A: TOC in Soil and Sediment

- USP <643> (for general TOC principles, if applicable)

- Instrument manufacturer’s manual (e.g., Shimadzu TOC-L with SSM-5000A)


#### 4. Definitions

- **TOC**: Total Organic Carbon, carbon bound in organic compounds.

- **TC**: Total Carbon, sum of organic and inorganic carbon.

- **IC**: Inorganic Carbon, carbon in inorganic compounds (e.g., carbonates).

- **TOC = TC - IC**: TOC is calculated by subtracting IC from TC.

- **Control Chart**: A graphical tool to monitor instrument performance over time.


#### 5. Equipment and Materials

- TOC analyzer with solid sample module (e.g., Shimadzu SSM-5000A)

- Analytical balance (0.1 mg precision)

- Ceramic or quartz boats (pre-combusted at 900°C)

- Drying oven (105°C)

- Desiccator

- Mortar and pestle or ball mill (for homogenization)

- Reagents:

  - Sucrose (for TC calibration and QC standards, 42.1% carbon by weight)

  - Calcium carbonate (CaCO₃, for IC standards, 12% carbon by weight)

  - Phosphoric acid (5–10% v/v, for IC removal if needed)

- Quality control standards (e.g., certified soil reference material or sucrose)

- Ultra-pure water (TOC < 50 ppb)

- Spatula and forceps (cleaned)

- Computer with data acquisition software


#### 6. Safety Precautions

- Wear PPE (gloves, lab coat, safety glasses).

- Handle acids (e.g., phosphoric acid) in a fume hood.

- Use caution with high-temperature furnaces (up to 900°C).

- Dispose of chemical waste per local regulations.


#### 7. Procedure


##### 7.1 Sample Preparation

1. **Collect and store samples**:

   - Collect representative samples in clean, airtight containers.

   - Store at 4°C and analyze within 28 days (or per regulatory guidelines).

2. **Dry samples**:

   - Air-dry or oven-dry at 105°C to constant weight to remove moisture.

   - Cool in a desiccator.

3. **Homogenize**:

   - Grind samples using a mortar and pestle or ball mill to a fine powder (<0.2 mm particle size).

   - Sieve to ensure uniformity.

4. **Remove IC (optional)**:

   - If high carbonate content is suspected, treat a subsample with 5–10% phosphoric acid to remove IC (add dropwise until fizzing stops, then dry).

   - Use this subsample for direct TOC measurement (TOC = TC).


##### 7.2 Instrument Preparation

1. **Turn on the TOC analyzer** and solid sample module.

2. **Set furnace temperatures**:

   - TC combustion: ~900°C.

   - IC reaction: ~200°C (for IC measurement).

3. **Check gases**:

   - Ensure carrier gas (e.g., high-purity oxygen or nitrogen) flows at ~500 mL/min.

4. **Clean boats**:

   - Pre-combust ceramic/quartz boats at 900°C for 10 minutes to remove residual carbon.

   - Store in a desiccator.

5. **Perform system blank**:

   - Run an empty boat to ensure TC < 0.1 µg C (or manufacturer’s limit).


##### 7.3 Calibration Procedure

1. **Prepare calibration standards**:

   - **TC standards**: Use sucrose (42.1% carbon by weight).

     - Weigh 1.19, 2.38, 4.75, 11.88 mg sucrose to yield 0.5, 1.0, 2.0, 5.0 mg C, respectively.

   - **IC standards**: Use calcium carbonate (12% carbon by weight).

     - Weigh 4.17, 8.33, 16.67, 41.67 mg CaCO₃ to yield 0.5, 1.0, 2.0, 5.0 mg C, respectively.

   - Place standards in pre-combusted boats.

2. **Run calibration**:

   - Analyze each standard in triplicate.

   - Record peak areas for TC (at 900°C) and IC (at 200°C with acid reaction).

   - Generate calibration curves:

     - Plot peak area vs. carbon mass (µg C) for TC and IC separately.

     - Use linear regression: y = mx + b (y = peak area, x = µg C, m = slope, b = intercept).

3. **Calibration criteria**:

   - **Correlation coefficient (R²)**: ≥ 0.995 for both TC and IC curves.

   - **Slope**: Must be consistent with instrument sensitivity (e.g., within ±10% of historical values).

   - **Intercept**: Near zero (e.g., < 5% of lowest standard’s peak area).

   - **Relative standard deviation (RSD)**: ≤ 5% for triplicate injections of each standard.

   - **Calibration check standard**: Run a mid-range standard (e.g., 2 mg C sucrose) after calibration. Recovery must be 95–105% of expected value.

4. **Re-calibration**:

   - Perform if any criterion fails or after significant instrument maintenance.


##### 7.4 Sample Analysis

1. **Weigh samples**:

   - Weigh 10–100 mg of dried, homogenized sample into a clean boat (adjust based on expected carbon content).

   - Record exact weight to 0.1 mg.

2. **Run sequence**:

   - **Blank**: Empty boat to confirm TC < 0.1 µg C.

   - **Calibration check standard**: Sucrose (e.g., 2 mg C) to verify calibration (95–105% recovery).

   - **Quality control sample**: Certified reference material or sucrose (e.g., 2 mg C).

   - **Samples**: Analyze in duplicate or triplicate.

   - **TC measurement**: Combust at 900°C.

   - **IC measurement**: React with acid at 200°C (if needed).

3. **Optional direct TOC**:

   - If IC was removed by acid treatment, measure TC of treated sample as TOC.


##### 7.5 Quality Control

1. **System Blank**:

   - Criteria: TC < 0.1 µg C (or manufacturer’s limit).

2. **Quality Control Sample**:

   - Use a certified reference material (e.g., soil with known TOC) or sucrose (e.g., 2 mg C).

   - Run every 10 samples or at least once per batch.

   - **Criteria**:

     - Recovery: 90–110% of certified/expected value.

     - Precision: RSD ≤ 5% for triplicate measurements.

     - If QC fails, investigate (e.g., check sample preparation, re-calibrate) and re-run.

3. **Control Chart**:

   - Plot QC results (%C or mg C) over time.

   - Calculate mean and ±2 standard deviations (control limits) from at least 20 historical QC results.

   - **Criteria**:

     - Results must fall within control limits.

     - No trends (e.g., 7 consecutive points above/below mean or increasing/decreasing).

     - Investigate and take corrective action if criteria are not met.

4. **Duplicates**:

   - Relative percent difference (RPD) between duplicates ≤ 10%.

   - RPD = |(C1 - C2)| / [(C1 + C2)/2] × 100.


##### 7.6 Computation for Obtaining TOC

1. **Determine TC and IC**:

   - Use calibration curves to convert peak areas to carbon mass (µg C).

   - TC (µg C) = (TC peak area - TC intercept) / TC slope.

   - IC (µg C) = (IC peak area - IC intercept) / IC slope.

2. **Calculate TOC**:

   - TOC (µg C) = TC (µg C) - IC (µg C).

3. **Convert to %C or mg C/g**:

   - %TOC = (TOC in µg C / sample weight in µg) × 100.

   - mg C/g = (TOC in µg C / sample weight in mg).

4. **Example Calculation**:

   - Sample weight: 50 mg (50,000 µg).

   - TC calibration: y = 1000x + 50.

   - IC calibration: y = 950x + 30.

   - TC peak area: 4500 → TC = (4500 - 50) / 1000 = 4.45 µg C.

   - IC peak area: 1000 → IC = (1000 - 30) / 950 = 1.02 µg C.

   - TOC = 4.45 - 1.02 = 3.43 µg C.

   - %TOC = (3.43 / 50,000) × 100 = 0.00686% (68.6 ppm).

   - mg C/g = 3.43 / 50 = 0.0686 mg C/g.


##### 7.7 Data Recording

1. Record sample weights, TC, IC, and TOC (µg C, %C, or mg C/g).

2. Save calibration curves, QC data, and control charts.

3. Document deviations, corrective actions, and instrument maintenance.


#### 8. Maintenance

- Clean combustion tube and boats after 100–200 runs or if contamination is suspected.

- Check gas regulators and replace if pressure drops.

- Perform annual maintenance by a certified technician.


#### 9. Troubleshooting

- **High blank values**: Re-combust boats or check carrier gas purity.

- **Calibration failure**: Verify standard weights, re-prepare standards, or clean furnace.

- **QC failure**: Check sample homogeneity, re-calibrate, or investigate contamination.

- **Inconsistent duplicates**: Ensure uniform grinding or check boat placement.


#### 10. Example

**Scenario**: Analyzing TOC in agricultural soil.

- **Sample Preparation**:

  - 100 g soil dried at 105°C, ground to <0.2 mm.

  - Sample weight: 50 mg.

- **Calibration**:

  - TC standards: 0.5, 1.0, 2.0, 5.0 mg C (sucrose).

  - IC standards: 0.5, 1.0, 2.0, 5.0 mg C (CaCO₃).

  - TC curve: y = 1000x + 50, R² = 0.998, RSD ≤ 3%.

  - IC curve: y = 950x + 30, R² = 0.997, RSD ≤ 4%.

  - Calibration check (2 mg C sucrose): 2.02 mg C (101% recovery, acceptable).

- **Analysis**:

  - Blank: 0.05 µg C (acceptable).

  - QC (2 mg C sucrose): 2.1 mg C (105% recovery, RSD = 2%, acceptable).

  - Sample TC: 4.5 µg C, IC: 1.0 µg C.

  - TOC = 4.5 - 1.0 = 3.5 µg C.

  - %TOC = (3.5 / 50,000) × 100 = 0.007% (70 ppm).

  - mg C/g = 3.5 / 50 = 0.07 mg C/g.

- **Control Chart**:

  - QC results (last 20 runs): Mean = 2.05 mg C, SD = 0.06 mg C.

  - Control limits: 1.93–2.17 mg C.

  - Current QC (2.1 mg C) within limits, no trends.


#### 11. Records

- Maintain a logbook for sample weights, calibration, QC, and results.

- Archive data for at least 5 years (or per regulatory requirements).


#### 12. Approval

- Prepared by: [Analyst Name]

- Approved by: [Lab Manager Name]

- Effective Date: [Date]


---


This SOP provides a comprehensive framework for TOC analysis in solid samples, with explicit calibration criteria, QC sample criteria, and TOC computation. Let me know if you need further customization, additional details, or clarification!

Comments

Popular posts from this blog

TOC analyser principle

1. A Total Organic Carbon (TOC) analyzer measures the amount of carbon in organic compounds present in a sample. The principle behind a TOC analyzer is based on the combustion or oxidation of organic materials to convert carbon into carbon dioxide (CO2), which is then measured to determine the total amount of organic carbon in the sample. There are two main methods for measuring TOC: Non-Combustion Method (Oxidation Method) : The sample is oxidized, usually using a strong oxidizing agent like persulfate or UV light, to convert organic carbon into CO2. The resulting CO2 is then measured using a non-dispersive infrared (NDIR) detector or another detection method to quantify the total organic carbon. Combustion Method : The sample is heated to high temperatures (around 680–950°C), where organic carbon is combusted in the presence of oxygen to form CO2. The CO2 produced is then measured using infrared spectroscopy or other appropriate detection techniques. Both methods al...

TDS- Total Dissolve Solid

Total Dissolved Solids (TDS) testing for wastewater is a critical procedure outlined in the Standard Methods for the Examination of Water and Wastewater by the American Public Health Association (APHA). Specifically, the method referenced is APHA Standard Method 2540C: Total Dissolved Solids Dried at 180°C. Below is a detailed Standard Operating Procedure (SOP) for preparing and conducting TDS testing in wastewater, including a description of the method and a flow chart to illustrate the process. Standard Operating Procedure (SOP) for TDS Testing in Wastewater (APHA Method 2540C) 1. Purpose To determine the concentration of total dissolved solids (TDS) in wastewater samples using the gravimetric method as specified in APHA Standard Method 2540C. TDS measures the mass of organic and inorganic substances dissolved in water that pass through a filter, typically after evaporation at 180°C. 2. Scope This SOP applies to the analysis of wastewater samples (industrial, municipal, or environmen...

Standard Operating Procedure for the Measurement of Total Hardness

  SOP For Total Hardness in Water and waste water EDTA Titrimetric method Purpose: The purpose of this protocol is to provide guidelines for analysis of Total Hardness in Water. Principle of method: EDTA and it's sodium salts from a chelated solution complex when added to a solution of certain metal actions, if a small amount of dye such as Eriochrome Black T is added to an aqoeous solution containing calcium and magnesium ions at a pH of 10+0.1 to 10-0.1, the solution becomes winr red. When titrated with EDTA , the magnesium and calcium will be complexed, and when all of the  and  has been complexed the solution turns from wine red to Blue.      To   Instrument/Equipment: The following items are necessary to perform the analysis of Total Hardness in Water; pH Meter Burette – 50 ml capacity Glass Wares Reagents/Chemicals: All the chemicals should be of Analytical Reagent Grade. Distilled Water Buffer Solution – Dissolve 16.9 g of Ammonium Chloride...